← All Advisories

Dräger SC Monitoring devices contain hard-coded credentials and denial-of-service vulnerability across all software versions

Last refreshed2026-09-30

Status: NEW  |  Advisory ID: CVE-2019-25722

Key Details

CVECVE-2019-25722
CVSS Score / Version7.6 (High) / CVSS v3.1
Updated2026-07-22
CVSS VectorCVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:L/I:L/A:H
CVSS Proseattack vector is adjacent; attack complexity is low; privileges required is none; user interaction is none; scope is unchanged; confidentiality impact is low; integrity impact is low; availability impact is high.
Classified asCWE-798 (Use of Hard-coded Credentials)

What to Know

Dräger SC Monitoring devices (SC 6002XL, SC 6802XL, SC 7000, SC 8000, SC 9000 XL) contain hard-coded plaintext credentials in source code and a denial-of-service vulnerability that allows local and remote attackers to compromise device integrity across all software versions. A local attacker with direct device access can use the hard-coded credentials to access service and clinical accounts and alter device configuration, while a remote attacker can send malformed network packets to cause repeated device reboots, ultimately resulting in loss of network connectivity and disruption of patient monitoring. (NVD)

References

SourceReference
NVDhttps://nvd.nist.gov/vuln/detail/CVE-2019-25722
CVEhttps://www.cve.org/CVERecord?id=CVE-2019-25722